Prostar XL Percutaneous Vascular Surgical System
VASCULAR
hamburger

SECURE

  • Uses two braided polyester sutures and four nitonol needles to provide a secure closure   
  • Vascular access can be maintained throughout deployment  

CLOSE 

  • Percutaneous vascular closure device  
  • Integrated pre-dilator to facilitate device placement 

IMPROVED OUTCOMES1    

  • Reduces time to hemostasis, ambulation and discharge   
  • Improves patient comfort and reduces access site complications.

1. Nasu et al. “The Clinical Effectiveness of the Prostar® XL Suture-Mediated Percutaneous Vascular Closure Device for Achievement of Hemostasis”; JIC, 2003; Vol 15, Issue 7

Prostar Close Access Sites

Prostar Secure Closure

Prostar Improved Outcomes

Prostar Closure System Information

*Improved outcomes as shown as compared to manual compression.

Designed for use in 8.5-10F femoral artery access sites following catheterization procedures.

Product Code/ Catalog # 12322-01

Sold with 5 systems per box

Included in Prostar XL System Package: 1 Prostar XL Device, 1 Knot Pusher

PRODUCT INFO

Our online product catalogs provide the very latest information available, right at your fingertips. Download catalogs by product family in PDF format, or view the entire Abbott Vascular inventory in Excel. Product barcodes enable direct scanning for easy reordering.

Indications And Important Safety Information

rx-only-logo Prostar® XL  Percutaneous Vascular Surgical System

INDICATIONS FOR USE

The Prostar XL PVS System is indicated for the percutaneous delivery of sutures for closing the common femoral artery access site and reducing the time to hemostasis and time to ambulation (patient walks ten feet) of patients who have undergone catheterization procedures using 8.5F to 10F sheaths. (Refer to PRECAUTIONS, SPECIAL PATIENT POPULATIONS sections).

CONTRAINDICATIONS

None known.

WARNINGS

The outer pouch of the Prostar XL PVS System and the individual accessories provides the sterile barrier.
Do not use the Prostar PVS System or accessories if the packaging or sterile barrier have been previously opened or damaged, or if the components appear to be damaged or defective.
DO NOT RESTERILIZE OR REUSE. The Prostar XL PVS System and accessories are intended for single use only. Do not use the Prostar XL PVS System if the puncture site is proximal to the inguinal ligament as this may result in a retroperitoneal hematoma.

PRECAUTIONS

1. The Prostar XL PVS device and accessories should only be used by physicians (or other healthcare professionals authorized by or under the direction of such physicians) after they have been trained in the use of the Prostar XL PVS System and accessories, e.g., participation in a Prostar XL PVS System training program or equivalent.

2. Observe sterile technique at all times when using the Prostar XL PVS System. Employ appropriate groin management, as per hospital protocol, post procedure and post hospital discharge to prevent infection.

3. Use a single wall puncture technique. Do not puncture the posterior wall of the artery.

4. Adequate knot security requires accepted surgical technique as warranted by surgical circumstances and the experience of the operator.

5. There are no reaccess restrictions if previous arteriotomy repairs were achieved with an Abbott Vascular Suture Mediated Device.

6. Do not insert the Prostar XL device into the femoral artery at an angle greater than 45 degrees to the longitudinal plane of the artery.

7. Do not advance or withdraw the Prostar XL device against resistance until the cause of that resistance has been determined (see CLINICAL PROCEDURE-Device Placement section). Excessive force used to advance or torque the Prostar XL device should be avoided as it may lead to significant arterial damage and / or breakage of the device, which may necessitate intervention and / or surgical removal of the device and arterial repair.

8. If excessive resistance in advancing the Prostar XL device is encountered, withdraw the Prostar XL device over a 0.038” (or smaller) guide wire and reinsert the introducer sheath or use conventional compression therapy.

9. In the event suture breakage occurs after an initial knot has been tied, care should be taken to avoid excessive force if the reintroduction of the Prostar XL device or introducer sheath is required. Any resistance to introduction should result in advancement of an introducer sheath small enough to be introduced without undue force.

10. If significant blood flow is evident through or around the barrel of the Prostar XL device, do not deploy needles. Remove the Prostar XL device over a 0.038” (or smaller) guide wire and insert an appropriately sized introducer sheath.

11. Remove the Prostar XL sheath before tightening the suture. Failure to remove the sheath prior to tightening the suture may result in detachment of the tip of the sheath.

12. Do not attempt to re-deploy Prostar XL needles after the needles have been “backed-down” into the sheath (refer to the TECHNIQUE FOR NEEDLE BACK-DOWN section).

13. In the event bleeding from the femoral access site persists after the use of the Prostar XL device and accessories, use conventional compression therapy.

ADVERSE EVENTS

The following adverse events have been reported and may occur include • Device Malfunction • Device Complication • Vascular Repair • Ultrasound Guided Compression • Transfusion • Infection Requiring IV Antibiotics • Hematoma > 6 cm • AV Fistula • Nerve Injury • Pseudoaneurysm

Prior to use, please reference the Instructions for Use at eifu.abbottvascular.com for more information on indications, contraindications, warnings, precautions, and adverse events.

AP2943673-WBU Rev. A

You are about to exit the Abbott family of websites for a 3rd party website

Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott.


The website that you have requested also may not be optimized for your screen size.

Do you wish to continue and exit this website?

false
accessibility
© 2016 Abbott. All Rights Reserved. Please read the Legal Notice for further details.

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

accessibility

You are about to exit the Abbott family of websites for a 3rd party website

Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott.


The website that you have requested also may not be optimized for your screen size.

Do you wish to continue and exit this website?